Fluke 5730A Calibration Manual page 89

Multifunction
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

Product
Frequency
Applied
2 mA
10 Hz
2 mA
20 Hz
2 mA
40 Hz
2 mA
1 kHz
2 mA
5 kHz
2 mA
10 kHz
20 mA
10 Hz
20 mA
20 Hz
20 mA
40 Hz
20 mA
1 kHz
20 mA
5 kHz
20 mA
10 kHz
200 mA
10 Hz
200 mA
20 Hz
200 mA
40 Hz
200 mA
1 kHz
200 mA
5 kHz
200 mA
10 kHz
2 A
20 Hz
2 A
1 kHz
2 A
5 kHz
2 A
10 kHz
Table 17. AC Current 2 mA to 2 A Verification
A40B AC
A40B
Resistance
Shunt
(Ohms)
10 mA
10 mA
10 mA
10 mA
10 mA
10 mA
20 mA
20 mA
20 mA
20 mA
20 mA
20 mA
200 mA
200 mA
200 mA
200 mA
200 mA
200 mA
2 A
2 A
2 A
2 A
5790A
Calculated
5790A
Reading
Current
Range
(Volts)
(Amps)
220 mV
220 mV
220 mV
220 mV
220 mV
220 mV
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
2.2 V
Multifunction Calibrator
Full Verification
Limits
610 nA
400 nA
274 nA
274 nA
570 nA
3 μA
6.1 μA
4 μA
2.74 μA
2.74 μA
5.1 μA
28 μA
61 μA
40 μA
26.4 μA
26.4 μA
48 μA
232 μA
600 μA
600 μA
1.02 mA
14.2 mA
79

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents