Fluke 5730A Calibration Manual page 80

Multifunction
Hide thumbs Also See for 5730A:
Table of Contents

Advertisement

5730A
Calibration Manual
Product
Output
0 μA
10 μA
-10 μA
200 μA
-200 μA
0 mA
2 mA
-2 mA
0 mA
20 mA
-20 mA
0 mA
100 mA
200 mA
-200 mA
0 A
1 A
2 A
-2 A
0 A
3 A
-3 A
10 A
-10 A
200 μA
-200 μA
Current =
70
Table 12. DC Current Accuracy Verification
Product
DMM
Range
Range
220 μA
200 mV
220 μA
200 mV
220 μA
200 mV
220 μA
2 V
220 μA
2 V
2.2 mA
200 mV
2.2 mA
2 V
2.2 mA
2 V
22 mA
200 mV
22 mA
2 V
22 mA
2 V
220 mA
200 mV
220 mA
200 mV
220 mA
2 V
220 mA
2 V
2.2 A
200 mV
2.2 A
200 mV
2.2 A
2 V
2.2 A
2 V
11 A
200 mV
11 A
2 V
11 A
2 V
11 A
200 mV
11 A
200 mV
11 A
2 V
11 A
2 V
DMM
Voltage
Re
ading
Shunt
Value
Shunt
Shunt
Value
(Ohms)
10 kΩ
10 kΩ
10 kΩ
1 kΩ
1 kΩ
1 kΩ
100 Ω
100 Ω
100 Ω
10 Ω
10 Ω
10 Ω
1 Ω
1 Ω
1 Ω
1 Ω
0.1 Ω
0.1 Ω
0.1 Ω
5725A Current Verification
10 Ω
0.1 Ω
0.1 Ω
0.01 Ω
0.01 Ω
1000 Ω
1000 Ω
DMM
Calculated
Voltage
Current
Reading
(Amps)
(Volts)
Limits
±7 nA
±7.4 nA
±7.4 nA
±15.4 nA
±15.4 nA
±8 nA
±78 nA
±78 nA
±50 nA
±.75 μA
±.75 μA
±0.8 μA
±5.3 μA
±11.4 μA
±11.4 μA
±15 μA
±85 μA
±235 μA
±235 μA
±480 μA
±1.5 mA
±1.5 mA
±3.9 mA
±3.9 mA
±20 nA
±20 nA

Hide quick links: