10. Connect the test tool to the 5500A as shown in Figure 5-7.
Figure 5-7. Volt Gain Calibration Input Connections 500V
11. Set the 5500A to supply a DC voltage of 500V.
12. Set the 5500A to operate (OPR).
13. Press
Gain DMM (CL0814) and Gain DMM (CL0815) will be calibrated now.
14. Wait until the display shows calibration status
15. Set the 5500A to 0V (zero) and to Standby.
16. Continue at Section 5.6.6.

5.6.6 Volt Zero

Proceed as follows to do the Volt Zero calibration:
1. Press
2. Terminate Input A and Input B with the BB120 and a 50Ω or lower termination.
3. Press
4. Wait until the display shows
5. Remove the 50Ω terminations from the inputs.
6. Continue at Section 5.6.8. (For firmware version V01.00 continue at Section 5.6.7).

5.6.7 Zero Ohm (firmware V01.00 only)

Proceed as follows to do the Zero Ohm calibration:
1. Press
2. Make a short circuit between the Input A banana socket and the COM input .
3. Press
4. Wait until the display shows the calibration status
5. Remove the Input A to COM short.
6. Continue at Section 5.6.8.
to start the calibration.
to select calibration adjustment step
to start the zero calibration of all mV/d settings (CL0820...CL0835)
Volt Zero (CL 0835):READY
to select calibration adjustment step
to start the Ohm Zero calibration of all ranges (CL 0840...CL 0846).
Calibration Adjustment
Gain DMM (CL0815):READY
Volt Zero (CL 0820):IDLE
.
Zero Ohm (CL 0840):IDLE
Zero Ohm (CL 0846):READY
5
5.6 Final Calibration
ST8129.CGM
.
.
.