Reference Voltage Circuit
This circuit derives several reference voltages from the 1.23V main reference source.
The output of an amplifier in the P-ASIC supplies a current to the +1.23V reference
source V301 via R307. The +3.3V REFPWM2 voltage is used as reference for the
PWMB outputs of the D-ASIC on the Digital part.
The +1.23V REFP voltage is used as main reference source for the reference circuit.
This circuit consists of four amplifiers in the T-ASIC, external gain resistors, and filter
capacitors.
Amplifier 1 and connected resistors supply the REFPWM1 reference voltage. This
voltage is a reference for the PWMA outputs of the D-ASIC on the Digital section. It is
also used as reference voltage for the LCD supply on the LCD unit.
Amplifier 2 and connected resistors supply the -1.23V REFN reference voltage, used for
the trigger level voltages TRIGLEV1&2, the C-ASIC POS-A and POS-B voltages, and
the C-ASIC OFFSET-A and OFFSET-B voltages. REFN is also the input reference for
amplifiers 3 and 4.
3-24
REFPWM2
+3.3V
R307
+1.23V
V301
1.23V
REFP
R309
GAINPWM
R311
+3.3V
R312
REFPWM1
GNDREF
R308
GAINREFN
R306
-1.23V
REFN
R310
GAINADCB
R303
+0.1V
REFADCB
R302
GAINADCT
R301
+1.6V
R305
REFADCT
REFATT
Figure 3-12. Reference Voltage Section
73
REFP
72
+
-
71
62
+
1
56
-
55
57
+
2
63
-
64
+
3
54
-
53
+
4
52
-
51
8
P-ASIC
OQ0256
T-ASIC
OQ0257