10.1 Software modifications
Changes and improvements made to the test tool software (firmware) are identified by
incrementing the software version number. These changes are documented on a
supplemental change/errata sheet which, when applicable, is included with the manual.
To display the software version, proceed as follows:
1. Press
2. Press
Section 5).
3. Press
10.2 Hardware modifications
Changes and improvements made to the test tool hardware are identified by incrementing
the revision number of the Main PCA. The revision number is printed on a sticker, see
the example below. The sticker is placed on D-ASIC D471, on the Main PCA.
This example of the Main PCA revision number
sticker indicates revision 1.
The following revisions have been released:
Revision 09
Revision number of first deliveries.
Revision 10
Changes:
Physical size of C511 and C512 changed
Reason:
For production purposes
Servicing effects:
none; you can use the PN listed in Section 8.
Revision 11
Changes:
C556 has been changed from 18 nF into 15 nF
Reason:
For production purposes
Servicing effects:
none; you can use the PN listed in Section 8.
to open the USER OPTIONS menu.
to show the VERSION&CALIBRATION screen (see Figure 5.1 in
to return to normal mode.
Modifications