Pulse Adjust Input B; Gain Dmm (Gain Volt) - Fluke ScopeMeter 123 Service Manual

Industrial scopemeter
Hide thumbs Also See for ScopeMeter 123:
Table of Contents

Advertisement

5.6.4 Pulse Adjust Input B

Proceed as follows to do the Pulse Adjust Input A calibration:
1. Press
2. Connect the test tool to the 5500A as shown in Figure 5-5.
3. Set the 5500A to source a 1V, 1 MHz fast rising square wave (SCOPE output,
MODE edge) (rise time ≤ 1 ns, aberrations <2% pp).
4. Set the 5500A to operate (OPR).
5. Press
6. Wait until the display shows
7. When you are finished, set the 5500A to Standby.
8. Continue at Section 5.6.5.

5.6.5 Gain DMM (Gain Volt)

Dangerous voltages will be present on the calibration source
and connection cables during the following steps. Ensure that
the calibrator is in standby mode before making any connection
between the calibrator and the test tool.
Proceed as follows to do the Gain DMM calibration.
1. Press
2. Connect the test tool to the 5500A as shown in Figure 5-6.
to select calibration step
Figure 5-5. 5500A Scope Output to Input B
to start the calibration.
Pulse Adj B (CL 0660):READY
Warning
to select the first calibration step in Table 5-3.
Pulse Adj B (CL 0660):IDLE
.
Calibration Adjustment
5.6 Final Calibration
ST8005.CGM
5
5-11