7. Clean the display glass with a moist soft cloth if necessary. Install the display
assembly. Ensure that the display is secured correctly by the four alignment tabs in
the top case. It is secured correctly when it cannot be moved horizontally.
8. Install the keypad pressure plate. Press the plate firmly, and slide it under the four
plastic keeper tabs in the top case.
9. Install the main PCA unit, and re-attach the cables. Secure the flat cables in the
connectors with the connector latches. Keep the backlight wires twisted to
minimize interference voltages! Insert the shielding flap below the main PCA
shielding plate.
10. Put the bottom case and the top case together at the flat cable side, and hinge the
cases to each other. This ensures the keypad foil flat cable is folded correctly.
11. Install the battery pack, and the battery door, see figure 6-5.
Figure 6-4. Mounting the display shielding bracket
Figure 6-5. Battery pack installation
Disassembling the Test Tool
6.5 Reassembling the Test Tool