Contrast Calibration Adjustment - Fluke ScopeMeter 123 Service Manual

Industrial scopemeter
Hide thumbs Also See for ScopeMeter 123:
Table of Contents

Advertisement

123
Service Manual

5.4 Contrast Calibration Adjustment

After entering the Maintenance mode, the test tool display shows
Warming Up (CL 0200):IDLE (valid)
Do not press
Maintenance mode again.
Proceed as follows to adjust the maximum display darkness (CL0100), the default
contrast (CL0110) , and the maximum display brightness (CL0120).
1.
2. Press
3. Using
4.
5. Press
6. Using
7.
8. Press
9. Using
10. You can now :
5-6
now! If you did, turn the test tool off and on, and enter the
Press
a three times to select the first calibration step. The display shows:
Contrast (CL 0100) :MANUAL
CAL. The display will show a dark test pattern, see Figure 5-2
adjust the display to the maximum darkness, at which the test pattern
is only just visible.
Press
to select the default contrast calibration. The display shows:
Contrast (CL 0110) :MANUAL
CAL. The display shows the test pattern at default contrast.
set the display to optimal (becomes default) contrast.
Press
to select maximum brightness calibration. The display shows:
Contrast (CL 0120) :MANUAL
CAL. The display shows a bright test pattern.
adjust the display to the maximum brightness, at which the test
pattern is only just visible.
Exit, if only the Contrast had to be adjusted. Continue at Section 5.7.
OR
Do the complete calibration. Press
and continue at Section 5.5.
Figure 5-2. Display Test Pattern
.
to select the next step (Warming Up),

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents