2.5 Environmental

Environmental

Temperature
Operating
Storage
Humidity
Operating:
Storage:
Altitude
Operating
Storage
Vibration
Shock
Fungus Resistance
Salt Exposure
Electromagnetic Compatibility (EMC)
Emission
Immunity
Enclosure Protection
2-10
@0 to 10 °C (32 to 50 °F)
@10 to 30 °C (50 to 86 °F)
@30 to 40 °C (86 to 104 °F)
@40 to 50 °C (104 to 122 °F)
@-20 to 60 °C (-4 to 140 °F)
RS-232, optically isolated
supports Epson FX, LQ, and HP Deskjet
Laserjet
, and Postscript
Serial via PM9080 (optically isolated
RS232 adapter/cable, optional).
Parallel via PAC91 (optically isolated
print adapter cable, optional).
Dump and load settings and data.
Serial via PM9080 (optically isolated
RS232 adapter/cable, optional), using
SW90W (FlukeView software for
Windows).
MIL 28800E, Type 3, Class III, Style B
0 to 50 °C (32 to 122 °F)
-20 to 60 °C (-4 to 140 °F)
noncondensing
95%
75%
45%
noncondensing
4.5 km (15 000 feet)
Max. Input and Floating Voltage 600
Vrms up to 2 km, linearly derating to 400
Vrms @ 4.5 km
12 km (40 000 feet)
max. 3g
max. 30g
MIL28800E, Class 3, 3.7.7 & 4.5.6.1
MIL28800E, Class 3, 3.7.8.2 & 4.5.6.2.2.
Structural parts meet 48 hours 5% salt
solution test.
EN 50081-1 (1992): EN55022 and
EN60555-2
EN 50082-2(1992): IEC1000-4-2, -3, -4, -5
(see also Section 2.8, Tables 2-1 to 2-3)
IP51, ref: IEC529
,