Check for Error Messages
N9310A User's Guide
The signal generator has two categories of instrument
messages: system messages and error messages. A system
message is triggered by operation errors, for example, setting
conflicts or data input is out of the range of a parameter.
An error message mainly is triggered by hardware defects
and has an ERR annunciator displaying on the screen.
In condition of a operation error occurs, the signal generator
reports a message at the bottom of the screen, indicating an
operation error occurred and the instrument has corrected
the error. The signal generator also automatically clears the
system messages 30 seconds later.
In condition of hardware defects or system error occurs, the
signal generator reports a message at the bottom of the
screen and also has an ERR annunciator displaying.
Here are some tips for you to check the signal generator for
error messages.
1
Check the display to see if the ERR annunciator is displayed.
If it is, press
the queue. Please refer to
for detailed system messages descriptions.
2
When you have reviewed and resolved all of the error
messages, press
messages.
3
Cycle the power on the signal generator and then check
again if the ERR annunciator is still there.
4
If you can not resolve all error messages, please contact
Customer Contact Center for service.
>Error Info to review each error messages in
Utility
Chapter
> Error Info > Clear softkey to delete the
Utility
Getting Started
7, "Instrument Messages"
2
23