Agilent Technologies N9310A User Manual page 123

Rf signal generator
Hide thumbs Also See for N9310A:
Table of Contents

Advertisement

Amplitude Start
Amplitude Stop
N9310A User's Guide
:SWEep:AMPLitude:STARt <val> <unit>
:SWEep:AMPLitude:STARt?
This command sets the start amplitude in an amplitude
sweep. The start amplitude should always be lower than the
stop amplitude in an amplitude sweep.
The query returns the current start amplitude.
Range and unit
Valid unit
Minimum increment
*RST value
Returned data format
Front panel access
:SWEep:AMPLitude:STOP <val> <unit>
:SWEep:AMPLitude:STOP?
This command sets the stop amplitude in an amplitude
sweep. The stop amplitude should always be higher than the
start amplitude.
The query returns the current stop amplitude.
Range and unit
Valid unit
Minimum increment
*RST value
Returned data format
Front panel access
Subsystem Command Reference
–127 to +13 dBm
dBm
0.1 dB
–127.0 dBm
<val> <unit>
Sweep > Step Sweep > Ampl Start
–127 to +13 dBm
dBm
0.1 dB
13.0 dBm
<val> <unit>
Sweep > Step Sweep > Ampl Stop
6
117