FM Deviation
FM Rate
N9310A User's Guide
:PM:DEViation <val> <unit>
:PM:DEViation?
This command sets the FM deviation.
Choices for the variables <val> depend on FM rate.
The query returns the current FM deviation.
Range
Valid unit
Minimum increment
*RST value
Returned data format
Front panel access
:PM:RATE <val> <unit>
:PM:RATE?
This command sets the frequency of the internal FM source.
The query returns the current frequency of the internal FM
source.
Range
Minimum increment
*RST value
Returned data format
Front panel access
Subsystem Command Reference
0 to 10 rad (300 Hz < FM rate < 10 kHz)
0 to 5 rad (10 kHz < FM rate < 20 kHz)
rad
0.001 rad
0.000 rad
<val> <unit>
FM > FM Deviation
300 Hz to 80 kHz
0.1 Hz
1.0000 kHz
<val> <unit>
FM > FM Rate
6