Agilent Technologies N9310A User Manual page 146

Rf signal generator
Hide thumbs Also See for N9310A:
Table of Contents

Advertisement

6
Subsystem Command Reference
LF Output Frequency
LF Output Amplitude
140
:LFOutput:FREQuency <val> <unit>
:LFOutput:FREQuency?
This command sets the LF output frequency.
The query returns the current LF signal frequency.
Range
Valid unit
Resolution
*RST value
Returned data format
Front panel access
:LFOutput:AMPLitude <val> <unit>
:LFOutput:AMPLitude?
This command sets the LF output amplitude.
The query returns the current LF output amplitude.
Range
Valid unit
Resolution
*RST value
Returned data format
Front panel access
20 Hz to 80 kHz
kHz, Hz
0.1 Hz
1.0000 kHz
<val> <unit>
LF Out > LF Out Freq
0 to 3 V
V, mV
1 mV
500 mV
<val> <unit>
LF Out > LF Out Ampl