:OVLenable (Overload Event Enable Register)

Command Syntax
:OVLenable <enable_mask>
<enable_mask> ::= 16-bit integer
The overload enable mask is an integer representing an input as described
in the following table.
The :OVLenable command sets the mask in the Overload Event Enable
Register and enables the reporting of the Overload Event Register. If an
overvoltage is sensed on a 50
to 1 M
Operation Status Register.
You can set analog channel input impedance to 50
N O T E
you can also set external trigger input impedance to 50
Table 43 Overload Event Enable Register (OVL)
Bit
15-11
10
9
8
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
658)
input impedance. If enabled, such an event will set bit 11 in the
Description
---
External Trigger Fault
Channel 4 Fault
Channel 3 Fault
input, the input will automatically switch
. If there are only two analog channels,
.
When Set (1 = High = True), Enables:
(Not used.)
Event when fault occurs on External Trigger input.
Event when fault occurs on Channel 4 input.
Event when fault occurs on Channel 3 input.