External:units - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

:EXTernal:UNITs

Command Syntax
:EXTernal:UNITs <units>
<units> ::= {VOLT | AMPere}
The :EXTernal:UNITs command sets the measurement units for the probe
connected to the external trigger input. Select VOLT for a voltage probe
and select AMPere for a current probe. Measurement results, channel
sensitivity, and trigger level will reflect the measurement units you select.
Query Syntax
:EXTernal:UNITs?
The :CHANnel<n>:UNITs? query returns the current units setting for the
external trigger.
Return Format
<units><NL>
<units> ::= {VOLT | AMP}
See Also
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
(see
page
658)
"Introduction to :EXTernal Trigger Commands"
"Introduction to :TRIGger Commands"
":EXTernal:RANGe"
on page 226
":EXTernal:PROBe"
on page 222
":CHANnel<n>:UNITs"
on page 411
on page 206
Commands by Subsystem
on page 218
5
227

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents