Trigger:spi:clock:timeout - Agilent Technologies InfiniiVision 5000 Series Programmer's Manual

Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

5
Commands by Subsystem

:TRIGger:SPI:CLOCk:TIMeout

Command Syntax
:TRIGger:SPI:CLOCk:TIMeout <time_value>
<time_value> ::= time in seconds in NR1 format
The :TRIGger:SPI:CLOCk:TIMeout command sets the SPI signal clock
timeout resource in seconds from 500 ns to 10 s when the
:TRIGger:SPI:FRAMing command is set to TIMeout. The timer is used to
frame a signal by a clock timeout.
Query Syntax
:TRIGger:SPI:CLOCk:TIMeout?
The :TRIGger:SPI:CLOCk:TIMeout? query returns current SPI clock timeout
setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR1 format
See Also