5
Commands by Subsystem
:TRIGger:SPI:CLOCk:TIMeout
Command Syntax
:TRIGger:SPI:CLOCk:TIMeout <time_value>
<time_value> ::= time in seconds in NR1 format
The :TRIGger:SPI:CLOCk:TIMeout command sets the SPI signal clock
timeout resource in seconds from 500 ns to 10 s when the
:TRIGger:SPI:FRAMing command is set to TIMeout. The timer is used to
frame a signal by a clock timeout.
Query Syntax
:TRIGger:SPI:CLOCk:TIMeout?
The :TRIGger:SPI:CLOCk:TIMeout? query returns current SPI clock timeout
setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR1 format
See Also
•
•
•
•
472
(see
page
658)
"Introduction to :TRIGger Commands"
":TRIGger:SPI:CLOCk:SLOPe"
":TRIGger:SPI:SOURce:CLOCk"
":TRIGger:SPI:FRAMing"
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
on page 411
on page 471
on page 476
on page 473