Commands by Subsystem

:ACQuire:DAALias

Command Syntax
:ACQuire:DAALias <mode>
<mode> ::= {DISable | AUTO}
The :ACQuire:DAALias command sets the disable anti- alias mode of the
oscilloscope.
When set to DISable, anti- alias is always disabled. This is good for cases
where dithered data is not desired.
When set to AUTO, the oscilloscope turns off anti- alias control as needed.
Such cases are when the FFT or differentiate math functions are silent.
The :DIGitize command always turns off the anti- alias control as well.
Query Syntax

:ACQuire:DAALias?

The :ACQuire:DAALias? query returns the oscilloscope's current disable
anti- alias mode setting.
Return Format
<mode><NL>
<mode> ::= {DIS | AUTO}