Connecting the 16047E

Connecting to the 4294A

Follow these steps below to connect the 16047E to the 4294A.
Step 1. Set the 16047E test fixture to the test connectors on the front panel of the 4294A
by gradually coupling the four BNC connectors and fastening screws of the fixture
with the test connectors and accessory mounting holes of the instrument until they
come to complete contact.
Step 2. Fasten two of the four BNC connectors to the mating test connectors by gradually
turning the two BNC connectors' rotation levers until each pair of connectors are
securely connected.Be sure to align the grooves on both sides.
Step 3. Turn clockwise the fixture's two fastening screws together, so that the fixture is
secured to the instrument.
Step 4. Finally, secure the remaining two BNC connectors of the fixture by turning
clockwise their rotation levers.
Figure 1-1
Connecting 16047E to 4294A
The fastening screws are designed to provide the fixture with mechanical stability
NOTE
and a prolonged life. Removing these screws will not affect measurement values.
Chapter 1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Installation Guide
Connecting the 16047E
11