Summary of Contents for Agilent Technologies 16034G
Page 1
(217) 352-9330 | Click HERE Find the Keysight / Agilent 16034G at our website:...
Page 2
Caution Take care to avoid rough handling and never allow any mechanical shock to the Agilent 16034G, especially against the contact pins from the sides or any to the parts mounted on top of the fixture. 注意 注意 注意 注意...
Page 4
Agilent 16034G Test Fixture Operation Manual Fourth Edition Agilent Part No. 16034-90011 January 2001 Printed in: Japan Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 5
Such noncompliance would also violate safety standards of design, manufacture, and intended use of the instrument. The Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements. • Ground the instrument...
Page 6
To avoid the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to a Agilent Technologies Sales and Service Office for service and repair to ensure that safety features are maintained in operational condition.
Page 7
Instruments are warranted for a period of one year. Fixtures and adapters are warranted for a period of 90 days. During the warranty period, Agilent Technologies will, at its option, either repair or replace products that prove to be defective.
Page 8
Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Sales and Service Office. Addresses are provided at the back of this manual. Safety Symbol General definitions of safety symbols used on the instrument or in its manuals are listed below.
Overview Product Overview Product Overview The 16034G is designed for chip type components whose sizes range from 0201 to 1306. This test fixture can take measurements of the chip type L,C,R. Figure 1-1 Product Overview Chapter 1 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
16034G has been checked mechanically and electrically. The contents of the shipment should be as listed in Table 1-1. If the contents are incomplete, if there is mechanical damage or defect, notify the nearest Agilent Technologies office.
Overview Functions Functions Figure 1-2 16034G Parts Part FUNCTION Contact Pin Contact for DUT electrode. LOW side Contact Pin connected to a instrument’s L and HIGH side Contact Pin connected to a instrument’s H Knob For lateral adjustment of HIGH side Contact Pin.
Operation This chapter describes the proper methods for open and short correction and DUT measurement. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
The following procedure shows correction and measurement by the 16034G. CAUTION Take care to avoid rough handling and never allow any mechanical shock to the 16034G, especially against the contact pins from the sides or any to the parts mounted on top of the fixture.
Operation Performing Open and Short Correction Performing Short Correction The short correction procedure is as follows. 1. Push the HIGH side contact pin’s knob to the left to make firm contact with the LOW side contact pin (Figure 2-2). Tighten the contact pin locking screw to secure the HIGH side contact pin.
Operation Performing Open and Short Correction Performing Open Correction The open correction procedure is as follows. 1. Push the HIGH side contact pin so that the distance between the HIGH and the LOW contact pins matches the DUT’s width (Figure 2-3). It is recommended that you place the DUT on the measuring stage and precisely position the HIGH side contact pin to actual DUT width.
Operation DUT Measurement DUT Measurement Before performing DUT measurement, open and short correction should be done as described in the previous sections. If measurement frequency is over 3 MHz, perform load correction before the DUT measurement described later onwards. 1. Adjust the HIGH side contact pin so that the DUT is positioned on the center of the measuring stage and secure the contact pin with the contact pin locking screw.
Operation DUT Measurement NOTE Measurement values can vary depending on contact pressure when measuring ferrite inductors or multi-layer ceramic capacitors with high permittivity. When measuring this kind of device, removing the E-ring can weaken the spring pre-load. However, this technique may increase contact resistance and thus degrades the accuracy of D parameter measurements.
Specifications and Supplemental Performance Characteristics This chapter provides specifications and supplemental performance characteristics of the 16034G test fixture. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Specifications and Supplemental Performance Characteristics Supplemental Performance Characteristics Supplemental Performance Characteristics This section provides useful data on the 16034G. These supplemental performance characteristics should not be considered specifications. Frequency Range ≤ With OPEN/SHORT correction 3 MHz ≤ With OPEN/SHORT/LOAD correction...
Page 25
Specifications and Supplemental Performance Characteristics Supplemental Performance Characteristics NOTE D is not expressed as a percentage but as an absolute value. Rs (ESR) Measurement Additional error Rse[%] of the Rs measurement is calculated by additional error Ze [%] of |Z| measurement as follows. If Dx ≤...
Specifications and Supplemental Performance Characteristics Supplemental Performance Characteristics Contact Pressure The following data are supplemental performance characteristics for the spring that applies contact pressure. Spring constant ± 10 % 37 gf/mm Spring pre-load Approximately 400 g (without E-ring, approximately 20 g) Chapter 3 Artisan Technology Group - Quality Instrumentation ...