Measuring 3-terminal device

The 16047E allows you to measure a 3-terminal device shown in Figure 3-3.
Figure 3-3
3-terminal device
Use the guard terminal on the fixture to measure a 3-terminal device.Connect the
lead No. 3 to the guard terminal in order to measure only characteristics of Z
eliminating any effects from Z
1. Loosen the electrode securing screws of both sides and the screw of the guard
2. Insert the leads No. 1 or 2 into the electrodes, then fasten the screws. Insert the
Figure 3-4
Measuring 3-terminal device
3. Follow the instruction manual that came with your instrument to measure your
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
terminal.
lead No. 3 into the guard terminal, then secure it with the screw.