Supplemental Performance Characteristics

This section provides useful data on the 16047E. These supplemental performance
characteristics should not be considered specifications.

Additional Error

Additional errors are calculated as follows.
|Z| Measurement
Additional error Ze [%] of the |Z| measurement is calculated by substituting the
values in the table below into the following equation.
Ze [%] = ± {A + (Zs/Zx + Yo ×Zx) × 100}
where
A [%]
Yo [S]
Zs [Ω]
Zx [Ω]
Without extension cable
Applicable Instruments : 4294A
Yo
Zs
A
f
f > 15MHz
where f is the measurement frequency (MHz).
Without extension cable
Applicable Instruments : 4192A, 4194A, 4263B, 4268A, 4278A, 4279A, 4284A,
4285A, 4338B
Yo
Zs
A
where f is the measurement frequency (MHz).
Chapter 4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Specifications and Supplemental Performance Characteristics
Supplemental Performance Characteristics
Additional Error (Proportional Error)
Open Repeatability (Admittance)
Short Repeatability (Impedance)
Measured Value (Impedance)
µ
2 n + 10
× ( f / 100 ) [S]
2 m + 600 m × ( f / 100 ) [Ω]
15 MHz
0.2 % × ( f / 10 )
4 % × ( f / 100 ) [%]
µ
2 n + 10
× ( f / 100 ) [S]
2 m + 600 m × ( f / 100 ) [Ω]
0.2 % × ( f / 10 )
2
[%]
2
[%]
27