Agilent Technologies 16047E Operation And Service Manual page 32

Test fixture for axial lead components
Table of Contents

Advertisement

Specifications and Supplemental Performance Characteristics
Supplemental Performance Characteristics
4194A (extension cable 1m)
where f is the measurement frequency (MHz).
4285A, 4294A (extension cable 1m)
where f is the measurement frequency (MHz).
4285A, 4294A (extension cable 2m)
where f is the measurement frequency (MHz).
D Measurement
Additional error De of the D measurement is calculated by additional error Ze [%]
of |Z| measurement as follows.
If Dx ≤ 0.1:
If 0.1 < Dx ≤ 0.5:
where Dx is the measured value of D. It is necessary for Ze to be below 10 %.
NOTE
D is not expressed as a percentage but as an absolute value.
28
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Yo
4 n + 20
Zs
4 m + 1200 m × ( f / 100 ) [Ω]
A
0.6 % × ( f / 10 )
Yo
4 n + 20
Zs
4 m + 1200 m × ( f / 100 ) [Ω]
A
f
5 MHz
0.6 % × ( f / 10 )
f > 5MHz
8 % × ( f / 100 ) [%]
Yo
4 n + 20
Zs
4 m + 1200 m × ( f / 100 ) [Ω]
A
f
5 MHz
1 % × ( f / 10 )
f > 5MHz
8 % × ( f / 100 ) [%]
De = Ze / 100
De = ( Ze / 100) × (1 + Dx)
µ
× ( f / 100 ) [S]
2
[%]
µ
× ( f / 100 ) [S]
2
[%]
µ
× ( f / 100 ) [S]
2
[%]