Agilent Technologies 16047E Operation And Service Manual page 33

Test fixture for axial lead components
Table of Contents

Advertisement

Rs (ESR) Measurement
Additional error Rse[%] of the Rs measurement is calculated by additional error
Ze [%] of |Z| measurement as follows.
If Dx ≤ 0.1:
If 0.1 < Dx ≤ 0.5:
Dx is the measured value of D and is calculated as follows.
Chapter 4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Specifications and Supplemental Performance Characteristics
Supplemental Performance Characteristics
Rse [%] = Ze / Dx
Rse [%] = ( Ze / Dx) ×
Dx = 2 × π × f × Csx × Rsx,
where
f: measurement signal frequency
Csx: measured value of Cs
Rsx: measured value of Rs.
2
(
)
1
+
Dx
29