6.4.4

Procedure - simulating alarms

Complete the following steps to make the analyzer energize or de-energize an alarm relay.
Instruction Manual
Choose Configure/Setpoint.
Choose Alarm 1, Alarm 2, Alarm 3, or Alarm 4.
The screens summarizes the present configuration and setpoints. There are nine
items: Setpoint, Assign (S1 is sensor 1), Logic, Deadband, Interval time, On time, Recover time,
and Hold while active. The last four items describe the operation of the timer. Only four
items are shown at a time. To view the remaining items, scroll to the bottom of the
screen and continue scrolling.
To make a change, move the cursor to the desired line and press ENTER.
A screen appears in which the present setting can be edited.
Press ENTER to store the setting.
For an explanation of terms, see
To return to the main display, press MENU and then EXIT.
Section 6.4.1
and
Section
Programming the transmitter
6.4.2.