Emerson Rosemount MCL-220 Manual page 36

Monochloramine system with rosemount 1056 transmitter
Hide thumbs Also See for Rosemount MCL-220:
Table of Contents

Advertisement

Programming the transmitter
Table 6-1:
Default Settings
Item
7. Simulate
Alarms
1. Logic
2. Assignments
a. AL1 and AL2
b. AL3 and AL4
3. Deadband
4. Interval timer settings
a. interval time
b. on time
c. recovery time
Measurement
1. Monochloramine (sensor 1)
a. units
b. resolution
c. input filter
Temperature related settings
1. Units
2. Temperature compensation
Security code
1. Calibrate/Hold
2. Program/Display
Calibration - Analog Outputs
1. 4mA
2. 20 mA
28
(continued)
Choices
0.00 to 22.00 mA
high or low
monochloramine, temperature, fault,
interval timer
monochloramine, temperature, fault,
interval timer
0 to 9999
0.0 to 999.9 hr
0 to 999 sec
0 to 999 sec
ppm or mg/L
0.01 or 0.001
0 to 999 sec
°C or °F
automatic or manual
000 to 999
000 to 999
0.000 to 22.000 mA
0.000 to 22.000 mA
Default
12.00 mA
AL1 low, AL2, 3, 4, high
monochloramine (sensor 1)
temperature (sensor 1)
0
24.0 hr
10 sec
60 sec
ppm
0.001
5 sec
°C
automatic
000
000
4.000 mA
20.000 mA