Emerson Rosemount MCL-220 Manual page 29

Monochloramine system with rosemount 1056 transmitter
Hide thumbs Also See for Rosemount MCL-220:
Table of Contents

Advertisement

4.
5.
Instruction Manual
Press ENTER.
The Output Range screen appears. The screen shows the present values assigned to
output 1 (O1) and output 2 (O2). The screen also shows which sensors the outputs
are assigned to. S1 is sensor 1. The assignments shown are the defaults for the
MCL-220. Outputs are freely assignable under the Configure menu.
For practice, change the 20 mA settings for output 1 to 8.5 ppm.
a. Move the cursor to the O1 S1 20 mA: 10.00 line and press ENTER.
The screen below appears.
b. Use the navigation keys to change 10.00 to 8.5 ppm. Use Left and Right to
move from digit to digit. Use Up and Down to increase or decrease the numeral.
c. To move the decimal point, press Left or Right until the decimal point is
highlighted. Press Up to move the decimal point to the right. Press Down to
move to the left.
d. Press ENTER to store the setting.
The display returns to the summary screen. Note that the 20 mA setting for output 1
has changed to 8.50 ppm.
Display and operation
21