Configuring Security Settings; Purpose; Procedure - Configuring Security Settings - Emerson Rosemount MCL-220 Manual

Monochloramine system with rosemount 1056 transmitter
Hide thumbs Also See for Rosemount MCL-220:
Table of Contents

Advertisement

6.7

Configuring security settings

6.7.1

Purpose

This section describes how to set security codes. There are three levels of security.
1.
2.
3.
The security code is a three digit number. The table shows what happens when different
security codes (XXX and YYY) are assigned to Calibration/Hold and All. 000 means no security.
Calibration/Hold
000
XXX
XXX
000
6.7.2

Procedure - configuring security settings

Follow the steps below to set security codes.
1.
2.
3.
Instruction Manual
A user can view the default display and diagnostic screens only.
A user has access to the calibration and hold menus only.
A user has access to all menus.
All
XXX
YYY
000
000
Press MENU.
The main Menu screen appears.
Move the cursor to Program and press ENTER.
Scroll to the bottom of the screen and continue scrolling until Security is highlighted.
Press ENTER.
Programming the transmitter
What happens
User enters XXX and has access to all
menus.
User enters XXX and has access to Cal-
ibration and Hold menus only. User
enters YYY and has access to all me-
nus.
User needs no security code to have
access to all menus.
User needs no security code to have
access to all menus.