Calibration
3.
4.
5.
6.
7.
8.
9.
48
Press MENU.
The main Menu screen appears. The cursor is on Calibrate.
Press ENTER.
Choose the sensor you wish to calibrate.
Sensor 1 is the monochloramine sensor.
Choose Temperature.
Change the display to match the temperature read from the calibrated
thermometer. Press ENTER.
If the present temperature is more than 5 °C different from the value entered, an
error message appears.
To force the transmitter to accept the calibration, choose Yes. To repeat the
calibration, choose No.
For troubleshooting assistance, see
To return to the main display, press MENU and then EXIT.
Section
9.5.
Rosemount MCL