Maintenance

Correcting Leaks

When required
Tools required
Parts required
1 Remove the front cover.
2 Use tissue to dry the leak sensor area.
3 Observe the capillary connections and the flow cell area for leaks and
4 Replace the front cover.
Leak plane
Waste outlet
Leak sensor assembly
Figure 32
96
If a leakage has occurred in the flow cell area or at the capillary connections.
Tissue
Two 1/4 inch wrenches for capillary connections
None
correct, if required.
Drying the Leak Sensor
1200 Series Variable Wavelength Detector User Manual