Installing the Detector

Flow Connections to the Detector

Preparations
Parts required
When working with solvents please observe appropriate safety procedures (for
WA R N I N G
example, goggles, safety gloves and protective clothing) as described in the
material handling and safety data sheet supplied by the solvent vendor, especially
when using toxic or hazardous solvents.
The flow cell is shipped with a filling of isopropanol (also recommended when the
N O T E
instrument and/or flow cell is shipped to another location). This is to avoid breakage due to
subambient conditions.
1 Press the release buttons and remove the front cover to
have access to the flow cell area.
36
Detector is installed in the LC system.
Other modules
Parts from accessory kit, see
Two wrenches 1/4–5/16 inch for capillary connections
2 Remove the metal cover and install the flow cell. Tighten
the cell screws.
1200 Series Variable Wavelength Detector User Manual
"Accessory Kit Contents"
on page 29