Using The Cuvette Holder - Agilent Technologies G1314B VWD User Manual

Variable wavelength detector
Table of Contents

Advertisement

8
Maintenance
Preparation for this procedure are:
Remove the normal flow cell.
Have cuvette with standard available.
2
Unscrew the bracket.
Next steps:
Reset the lamp counter as described in the User Interface documentation.
Turn the lamp ON.
Give the lamp more than 10 minutes to warm-up.
Perform
"Wavelength Verification/Calibration"
94
1
Locate the cuvette holder on the desk.
3
Insert the cuvette with the sample into the holder. The
clear side of the cuvette must be visible.
Light path
on page 104to check the correct positioning of the lamp.
1200 Series Variable Wavelength Detector User Manual
Clear side