Technical data

Electrical Data
Mains voltage *
400 V ±10%, 50/60Hz
Mains socket
(3,N,PE) AC,
CEE (32A, 5pol.,6h)
Max. consumption
approx. 12KVA
Lamp power
2200VA
Max. lamp power
max. 2800VA
Rated current
16A
Lamp current
max. 20A
Equipment Data
Dimensions
900mm x 1200mm x 1800mm
width x depth x
height
Weight
approx. 400kg
Sample surface
max. 4000cm²
Comment:
* = Other voltages on request.
** = Max. values achievable are dependent on the chosen filter system, the irradiance settings a the laboratory
environmental conditions.
Operating manual XENOTEST BETA LM
Operational Data
Cooling air required for
max. 200m³/h
lamps
Cooling air required for
max. 200m³/h
test chamber
Ultra pure water for rain
max. 1l/min
Ultra pure water for
max. 4l/h
humidity
Ultra pure water for
max. 1l/min
sample back side
cooling (Option)
Specification
Irradiance at sample level in
UV range 300 - 400 nm
XENOCHROME 270
60 - 165 W/m²
XENOCHROME 300
45 - 120 W/m²
XENOCHROME 320
35 - 100 W/m²
Black standard
Ambient - 130°C
temperature **
Test chamber temperature
Ambient - 80°C
**
Rel. humidity **
10-95%