Trouble Shooting - Atlas XENOTEST BETA LM Operating Manual

Table of Contents

Advertisement

Operating manual XENOTEST BETA LM
8.8

Trouble shooting

In the XENOTEST BETA all weathering programs
and unit functions are subject to permanent control.
Input errors are indicated in the display of the pro-
gram control system. The measures indicated for
No.
Fault
1
Lamp too old or irradiance cannot be
reached with current filter combination
2
Irradiance can`t be reached.
3
BST can`t be reached
4
Chamber temperature can`t be reached • Selected black standard temperature too low,
Page 48
debugging can be carried out by the user in
compliance with the applicable safety regulations.
Otherwise, please call our ATLAS service
department.
Possible measures (Spare part no.)
• Cleaning of the filters with alcohol and a lintfree cloth,
• Check of the fitted filter system or outer cylinder,
• Check of the selected filter system in the weathering program,
• Check of the lamp age, if necessary exchange lamp
NXE 2200 (no. 56076400),
• Exchange the filters
• Maximum deviation of program parameters too small,
• Time interval until error message too short,
• Check of fan speed,
• Check XENOSENSIV sensor with Radialux and calibrate if
necessary,
• Cleaning of the filters with alcohol and a lintfree cloth,
• Check of the fitted filter system or outer cylinder,
• Check of the selected filter system in the weathering program,
• Check of the lamp age, if necessary exchange lamp NXE 2200 (no.
56076400),
• Exchange the filters.
• Maximum deviation of program parameters too small,
• Time interval until error message too short,
• Check of fan speed,
• Check XENOSENSIV sensor with Radialux and calibrate if
necessary,
• Check of the irradiance.