Operating manual XENOTEST BETA LM
8.
Program example XENOTEST BETA
8.1
Task of the program
The sample program is to run with the following parameters:
Parameter
Filter system
Program number
Number of phases
Irradiance (E)
Temperature control
Test chamber
temperature (CHT)
Rain
Sample backside
cooling
Relative humidity in %
Phase time
Switch off criterion
Total time
8.2
Solving the program task
DISPLAY TEXT
ATLAS
XENOTEST GmbH
Unittype __
VERSION x.y ---> ENTER
A)
Input of filter system:
BASIC PARAMETER
PROGRAM START
PROGRAMMING
MAINTENANCE
FILTER SYSTEM
CHANGE OF LAMP
CALIBRATION
Comment: Use the „DATA" key to retrieve the possible filter systems and confirm with „ENTER".
XENOTEST BETA
Safety hints
Program example on
Function
XENOTEST BETA
Page 40
XENOCHROME 300
1
2
Total time
24 Hours
= 1
= 2
= 3
= 4
= 1
= 2
= 3
---> DATA
Unit set-up
Features
Test set-up
Phase 1
80 W/m²
Test chamber
temperature
65 °C
No
No
40 Percent
102 min
Options and accessory
Unit description
of the XENOTEST BETA
Care and cleaning
ATLAS
Phase 2
80 W/m²
Test chamber
temperature
55 °C
Yes
No
Rain
18 min
INPUT/KEY PAD
ENTER
1
1
DATA
Start-up
List of options
Maintenance
and accessories