Operating manual XENOTEST BETA LM
7.1.4

Control of the test chamber temperature

The test chamber temperature is measured in the
air flow and then displayed. By adding air of the
laboratory and in conjunction with the heating
7.1.5
Control of the test chamber humidity
The test chamber humidity is produced by two
humidifiers (fig. 8.2). It is measured in the air flow
by a sensor and indicated in percent relative
humidity. Figure 30 shows the setting options for
the achievable relative air humidity depending on
Figure 30: Humidity diagram
100
Fan spe ed
90
800
1250
80
2000
2500
70
60
50
40
30
20
10
0
10
20
7.2

Rain

The rain sprinkling of the entire sample surface is
pressureless and is provided by three rain bars
placed in the test chamber.
XENOTEST BETA
Safety hints
Program example on
Function
XENOTEST BETA
Page 38
30
40
chamber tempera ture [°C]
Unit set-up
Features
Test set-up
aggregate the test chamber temperature can be
controlled in a range from ambient °C to 80°C.
the test chamber temperature. The diagram shows
the working fields as parameters for the various
fan speeds. The fan speed can be adjusted in a
range between 800 and 2500 rpm.
50
60
7.3

Sample back cooling

The sample back cooling comes from spray
nozzles provided in the test chamber wall.
Options and accessory
Unit description
of the XENOTEST BETA
Care and cleaning
Maintenance
ATLAS
70
80
Start-up
List of options
and accessories