3. Unit set-up
To ensure the functioning of the water circuit and
the humidification system, the apparatus
must be aligned by means of its height
adjustable feet.
3.1
Scope of delivery
XENOTEST BETA LM
The unit is supplied in a suitable shipping case.
Sensitive components such as:
1.
Xenon lamps,
2.
optical filter system,
3.
sample holders and cover masks,
4.
XENOSENSIV sensor,
5.
outer cylinder,
are packed separately.
The scope of delivery includes (other fittings are
available when wished):
1.
Control system with display and operating
keyboard,
2.
3 Xenon lamps (type NXE 2200),
3.
Sensor to measure test chamber
temperature and rel. humidity,
4.
XENOSENSIV sensor to measure the
irradiance in the UV range from 300-
400 nm and the black standard tem-
perature,
5.
End plug for operation without
XENOSENSIV sensor,
6.
Support sheet to fix the XENOSENSIV
sensor on to the sample holder,
7.
XENOCHROME filter system consisting of
filter lantern with 11 special filters,
8.
Outer cylinder made of quartz glass and
special UV glass (suprax),
9.
16 standard sample holders,
10.
16 masking sheets to fix unstable samples
on the sample holder,
11.
Water tank,
12.
Thermo printer
13.
Interface RS232 / RS485
14.
Technical documentation:
Operating manual, circuit diagrams, spare
parts list.
XENOTEST BETA
Safety hints
Program example on
Function
XENOTEST BETA
Unit set-up
Features
Test set-up
Operating manual XENOTEST BETA
3.2

Space requirements

3.2.1

Laboratory

To avoid contamination and to ensure reproducibi-
lity of the climatic conditions, we recom-
mend to install the equipment in an air-
conditioned and air-filtered (filter type EU4)
laboratory at 18°C - 25°C and a relative
air humidity of 50%.
3.2.2

Room ventilation

The XENOTEST BETA needs 200 m³/h air to cool
down the xenon lamps and 200 m³/h air to
cool down the test chamber. An amount
of 400 m
/h fresh air has to be supplied to
3
the laboratory via an air filter (filter type
EU4).
3.2.3
Electrical connection
The unit is designed for a mains supply of
400V ± 10%. Connection is ensured by
safety contact plug (3,N,PE), CEE (32,
5pol., 6h).
For the electrical supply observe UVV
VGB 4.
Mains conditions have to agree with the
data on the identity label .
Observe VDE and technical connection
regulations of the energy supply compa-
nies.