Option And Accessories Of Xenotest Beta; Options; Rain-, Humidity Function And Sample Cooling - Atlas XENOTEST BETA LM Operating Manual

Table of Contents

Advertisement

ATLAS
5. Options and accessories of the XENOTEST BETA
5.1

Options

Unit components which are installed by Xenotest GmbH.
5.1.1
Rain, humidity function and sample back cooling (fig. 11)
The water tank (1) is filled through a water inlet
socket (2) (see chapter 4.1.3 on water quality).
The overflow is regulated by a level switch (3),
water shortage is indicated by a solenoid switch
(4) on the display. Two diving pumps (5) are
located in the water tank which supply the two
humidifiers (6) with water. The supply amount is
Figure 11: Water circuits
7
6
5
2
3
4
1
1
Water tank
2
Water inlet socket
3
Level switch
4
Solenoid switch
5
Diving pump (2x)
6
Humidifier (2x)
7
Valve
XENOTEST BETA
Safety hints
Program example on
Function
XENOTEST BETA
Unit set-up
Features
Test set-up
Operating manual XENOTEST BETA LM
adjustable by a valve (7). The rain pump (8)
supplies water to the rain bars (9). The pump (10)
supplies water through the supply lines (12) to
the spray nozzles for the sample back cooling
(11) . Excess water is returned to the water tank
via a reverse hose (13) or evacuated outside. The
water tank can be emptied by the water outlet
socket (14).
11
9
12
8
13
10
14
8
Rain pump
9
Rain bar (3x)
10
Pump for sample back cooling
11
Spray nozzle for sample back cooling
12
Water inlet hose
13
Water reverse hose
14
Water outlet socket