Operating manual XENOTEST BETA LM
4.2
Right hand apparatus side
(fig.8)
The mains switch (1) (fig. 7) is located at the top
left; disconnect unit in case of trouble.
The unit must be disconnected for all maintenance
works. Turn switch to position "0", and pull
mains plug. Protect against replugging!
Figure 7: Main switch
0
A thermoprinter (2) is placed above the main
switch providing for the print-out of data of the stored
programs as well as of the actual measuring values
of the current program.
The interface (3) on the right-hand side of the main
switch provides for the data output and external
computer control.
Right to the test chamber (4) is the climatic duct
(5) with the integrated heating (6) to warm up the
test chamber air. The two air flap motors (7) serve
to adjust the air flaps in the ventilation duct (8).
The climatic fan (9) recirculates the air and is
controlled by the climatic fan control (10).
The operating hour counter (11) counts the
operating time of the equipment.
The three power supply units (13) of the lamps
are placed behind a protective wire mesh (12).
XENOTEST BETA
Safety hints
Program example on
Function
XENOTEST BETA
Page 18
1
Unit set-up
Features
Test set-up
High voltage is present on the power packs.
Touching may be hazardous to your life !
The fuse for the heating (14) and the fuses for
the three lamps (15) are located below the
protective wire mesh. The fuse protection for the
three lamps is 25 amps and for the heating it is 16
amps.
4.3
Apparatus top (fig. 8)
The separate exhaust air flows for lamp cooling and
test chamber warming are evacuated at the top of
the unit via separate exhaust air sockets to the open
(see fig. 11 ventilation scheme). Exhaust air sok-
ket lamp cooling (16) diameter 100mm, exhaust
air socket test chamber warming (17) diameter
200mm.
The hoses for the lamp and test chamber
exhaust air must be made of flexible heat-
resisting material, as the temperatures of
the exhaust air may reach up to a max. of
120°C.
The max. length of the exhaust hose is li-
mited to 3m including the option of two 90
degree bends. In case of longer exhaust
air ducts or with more than two 90 degree
bends an external blower is needed.