National Instruments VXI-MXI-Express Series User Manual page 62

Hide thumbs Also See for VXI-MXI-Express Series:
Table of Contents

Advertisement

Configuring the Wired Trigger and Clock Bus
© National Instruments
Each of the 8 physical trigger channels is independently configurable and
designed to be used in a point-to-multipoint topology where one device acts
as the trigger driver and the remaining devices are listeners.
To map a trigger from the VXI backplane to multiple chassis, configure the
trigger to be mapped as Out of the Frame in MAX. This configures the
trigger as a driver. The remaining devices on the wired VXI backplane
trigger and CLK10 bus need to be configured as listeners by configuring
their respective triggers as In to the Frame.
To map CLK10 Out of the Frame set jumper W7 to connect pins 1-2. To
map CLK10 In to the Frame set jumper W7 to connect pins 2-3. Refer to
Appendix B,
Advanced Hardware Configuration
information about setting the hardware jumper.
Appendix C
Using the Trigger Ports on the NI VXI-8360T
C-3
Settings, for more
VXI-MXI-Express Series User Manual