7.1.3 Input Calibration Procedure

7.1.3.1 Correct Sensor 1 / 2 Input at Low Calibration Point
1. After Low and High Calibration points have been entered, as described above, select the S 1 /
2 cal lo Corrects button on the CALIBRATION display.
2. Select the S 1 / 2 cal lo Corrects button. This message appears:
3. Check the device is not in write protect mode and press OK
4. Enter calibration date (MM/DD/YY), OK
5. Enter calibration Time (Hour), OK. Entre calibration time (minute), OK
6. Apply Input on Sensor 1.
7. Adjust the PV input temperature to the exact value of the Lower calibration value entered in
the CONFIGURE display. Select the OK button.