Input Calibration Procedure; Correct Input At The Lower Range Value (Lrv); Figure 11 - Input Calibration Connections - Honeywell SmartLine STT700 HART User Manual

Stt700 series hart/de option
Hide thumbs Also See for SmartLine STT700 HART:
Table of Contents

Advertisement

6.8 Input Calibration Procedure

6.8.1 Correct Input at the Lower Range Value (LRV)

This section describes the action to correct the input at the Lower Range Value (LRV). To begin,
enter the appropriate values before doing the calibration. Configure sensor type to mV or TC or RTD
or Ohms for the input which you are going to perform calibration. Below snapshots are examples of
sensor type TC-E.
These values are entered in Device Setup/Sensors/Sensor 1/Sensor 1 Config Params menus, thus the
first set is for LRV1 Correct and URV1 Correct and the second set is for LRV2 Correct and URV2
Correct.
1. Select the Correct LRV button on the CALIBRATION display.
2. Select the Correct LRV button. This message appears:
Page 70
Figure 11 – Input Calibration Connections
STT700 Series HART/DE Option User's Manual
Revision 1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Smartline stt700 de

Table of Contents