Table 12 - Calibration Menus - Honeywell SmartLine ST 800 User Manual

Pressure transmitters
Table of Contents

Advertisement

<Return> Return to the Level 1 menu
<Return>
Hour
Minute
Year
Set Time
Month
Stamp
Day
<Return>
Zero
Correct
Do Zero Correct
<Return>
LRV
Correct
Do LRV Correct
<Return>
URV
Correct
Do URV Correct
<Return>
Cal Low Point
Cal High Point
Calibration
Parameters
Cal Adjust Ratio
Barometric
Standard (Available
for AP models
only)
33
Table 12 – Calibration Menus
# #
# #
# # # #
January -
December
# #
Executing this selection corrects the Zero
based on the input pressure.
The current live value of the primary pressure input
is shown on this display so the user can easily see
the effect of the Zero correction.
Executing this selection corrects the LRV
based on the input pressure.
The current live value of the primary pressure input
is shown on this display so the user can easily see
the effect of the LRV correction.
Executing this selection corrects the URV
based on the input pressure.
The current live value of the primary pressure input
is shown on this display so the user can easily see
the effect of the URV correction.
######.##
######.##
#.##
###.###
ST 800 Pressure Transmitter User's Manual
These selections allow the user to
enter a time stamp for the Zero
Correct, LRV Correct, URV
Correct, and Reset Corrects. This
time stamp can be read via HART
and FF communications.
The lower pressure value
for a two-point
field calibration within
the LRV to URV span.
The upper pressure value
for a two-point field
calibration within the
LRV to URV span.
A ratio value that adjusts
the calibrated curve
between the Cal
Low and High Points.
The pressure value for an
absolute pressure
correction when the
pressure input is open to
atmosphere.
Press  to enter menu
selection
 and  to select
number.
 to enter and shift to
next digit
 and  to select
from list.
 to enter
Press  to enter
menu selection
Scroll to Do Zero
Correct.
Press  to initiate
Press  to enter
menu selection Scroll
to Do LRV Correct.
Press  to initiate
Press  to enter
menu selection Scroll
to Do URV Correct.
Press  to initiate
Press  to enter
menu selection. Press
 and  to select
number. Press  to
enter and shift to
next digit.