Emerson Rosemount 5408 Reference Manual page 86

Level transmitters non-contacting radar
Hide thumbs Also See for Rosemount 5408:
Table of Contents

Advertisement

Configuration
3.
4.
5.
76
Set the desired transmitter variable to use for Primary Variable (PV), Secondary
Variable (SV), Third Variable (TV), and Fourth Variable (QV).
a. Select Configure > Manual Setup > Device Setup > HART.
b. Under Variable Mapping, select variables for PV, SV, TV, and QV.
Set the Rosemount 5408 to Burst Mode.
HART Revision 6:
a. Select Configure Burst Mode.
b. Under Burst Mode, select On.
c. Under Burst Command, select PV, SV, TV, QV.
d. Select Send.
HART Revision 7:
a. Select Configure Burst Mode.
b. Select View/Configure Message 1.
c. Under Message 1 Broadcast, select Wired HART Enabled.
d. Under Burst Command, select PV, SV, TV, QV, and then select Next.
e. Under Burst Msg Trigger Mode, select Continuous, and then select Next.
f. Set the Update Period, and then select Finish.
Prior to exiting the configuration, note the selected variables for SV, TV, and QV, and
the units set for each of the variables. The same configuration must be used for the
Rosemount 333.
Refer to the Rosemount 333 HART Tri-Loop
about installing and configuring the Rosemount 333.
Reference Manual
for full information
Reference Manual