Mechanical Installation
Procedure
1.
2.
3.
50
Mount the protection plates to flange/manhole cover. These plates protect the
internal surfaces of the flange ball from dust and sparks during welding.
G
E
F
E
Weld the flange ball.
Remove the protection plates and visually inspect the internal surfaces of the flange
ball for damage and dirt.
Ø 3.94 ± 0.02 in.
(Ø 100 ± 0.5 mm)
Max. 1.18 in. (30 mm)