Pt5 - Electronics Wrap Around - Teledyne WORKHORSE Operation Manual

Horizontal h-adcp
Hide thumbs Also See for WORKHORSE:
Table of Contents

Advertisement

May 2015
>pt4
IXMT
=
0.0 Amps rms
VXMT
=
19.3 Volts rms [Data=4ch]
Z
=
999.9 Ohms
Transmit Test Results = $C0 ... FAIL
>

PT5 - Electronics Wrap Around

The PT5 test is not a valid test and will be removed in a future firmware release.
This test sets up the H-ADCP in a test configuration in which the test output lines from the DSP Board
timing generator are routed directly to the Receiver board. The receiver then processes this signal. The
test output signal sends a certain correlation pattern when processed. The ideal pattern is as follows.
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
255 255 255 255
0
0
0
0
0
0
0
0
255 255 255 255
0
0
0
0
0
0
0
0
0
0
0
0
255 255 255 255
Acceptable deviations from this pattern are due to deviations in sampling bandwidth and demodulator
low-pass filter bandwidth variations. For example:
>PT5
13
13
13
13
13
13
13
13
13
13
13
13
0
0
0
0
255 255 255 255
13
13
13
13
0
0
0
0
255 255 255 255
13
13
13
13
13
13
13
13
0
0
0
0
255 255 255 255
Electronics Test Results = $0000
Page 162
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
[Data= 0h]
WorkHorse H-ADCP Operation Manual

Hide quick links:

Advertisement

Table of Contents
loading