Table 7: Pre-Deployment Test (Pa) Possible Cause Of Failures - Teledyne WORKHORSE Operation Manual

Horizontal h-adcp
Hide thumbs Also See for WORKHORSE:
Table of Contents

Advertisement

May 2015
Your H-ADCP fails any of the following PA tests provided the items indicated by {} have been
checked:
Recorder Tests:
Any recorder tests fails {provided that the PCMCIA card(s) have been checked for proper
installation, operation and they are DOS formatted; we STRONGLY recommend checking
PCMCIA cards in a computer before replacing the boards}
System Tests:
Transmit {if the H-ADCP fails when it is in water and air bubbles have been rubbed from
the faces}
Sensor Tests:
H/W Operation {if the H-ADCP fails when it is NOT sitting/resting on its side, or located
near a large magnetic field like a motor in a boat}
The spare boards kit will not correct any of the following failures:
A damaged beam or its urethane surface
Damage to the transducer beam connections below the copper shield
If it passes all PA tests and yet the data is all marked as bad
Fails the following PA test:
System Tests:
Wide Bandwidth {bandwidth tests may fail due to external interference}
Narrow Bandwidth {bandwidth tests may fail due to external interference}
RSSI Filter
Transmit
Table 7:
Pre-deployment Test (PA) Possible Cause of Failures
PA Test Name
Pre-Deployment Tests CPU Tests:
RTC
RAM
ROM
Recorder Tests:
PC Card #0
Card Detect
Communication
DOS Structure
Sector Test (short)
PC Card #1
Card Detect
Communication
DOS Structure
Sector Test (short)
DSP Tests:
Timing RAM
Demod
RAM
Demod
REG
FIFOs
System Tests:
XILINX Interrupts
Receive Loop-Back
Wide Bandwidth
Narrow Bandwidth
RSSI Filter
Transmit
Sensor Tests:
H/W Operation
Page 88
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Possible Cause of Failure
CPU board failed
PC card not plugged in
PC card failed
DSP board failed
DSP board failed
DSP or CPU board failed
DSP or CPU board failed
Not in water
External interference
DSP or Receiver board failed
Not in water or PIO board failed
PIO board failed
Receiver board failed
Pressure sensor failed
H-ADCP laying on its' side
WorkHorse H-ADCP Operation Manual