CableIQ Qualification Tester
Users Manual

Signal Performance Results

Note
Signal performance results for cables
less than 13 ft (4 m) long may be
unreliable.
Qualified/unqualified result for the signal
A
performance. Signal performance includes
crosstalk, insertion loss, and return loss. For
1000BASE-T, ELFEXT (equal-level far-end
crosstalk) is also tested.
28
B
Information about the signal performance:
Qualified for <application name>: The cabling will support the
application.
No Signal Performance tests needed for qualification: Appears
for the Telco and Wiremap only tests.
Wiring fault! Not tested.: The tester does not evaluate signal
avv23f.eps
performance if there is a wiremap, length, or delay skew fault.
Connection fault: A localized crosstalk fault was detected.
Localized faults are usually caused by bad connections. Check
the cabling at the location given. See the Appendix for other
causes of crosstalk faults.
Distributed cable faults: Crosstalk or an impedance problem was
detected along most or all of the cabling. The cabling is of poor
quality or is the wrong category for the selected application.
Figure 11. Signal Performance Results
-continued-