Autotest on Coaxial Cabling (75 Ω)
1
Turn on the tester and turn the rotary switch to
AUTOTEST.
2
Press
K c
. If the coax test is disabled, press
H
or
Setup to enable it.
J
3
Connect the tester and wiremap adapter or ID
locator to the cabling as shown in Figure 13.
P
4
Press
. The Autotest proceeds when a wiremap
adapter or ID locator is detected. The analog toner
turns on if neither is detected.
5
To view results, press
J M
32
. See Figure 14.
6
To save the results:
a.
Press
.
K S
AD
b.
Use
and
H
or Outlet field for editing. You can enter up to
17 characters in each field. See Figure 4 on page
13 for details on editing text.
To see a list of predefined labels for the
highlighted field, press
AD
, or SHIFT +
B C
H
list. Use
or
J
c.
On the Enter ID screen, press
the results with the Site, Location, and Outlet
entries shown.
The last character of the Outlet string
increments each time you save an Autotest.
to select the Site, Location,
AD
J
List. Use
to move through the
to select an entry.
K S
to save
Note
,