Technical Reference Handbook
Running the Test in Loopback Mode
W
Caution
If you disconnected the reference test cord from
the tester's output, you must set the reference
again to ensure valid measurements.
1
Verify that the settings described in Table 6-1 are
appropriate. Let the tester warm up for 5 minutes.
2
Select Loss/Length mode: On the HOME screen, press
A
Change Test; then select Loss/Length.
3
Clean the connectors on the cabling to be tested.
4
Connect the tester to the cabling as shown in Figure
6-10.
6-28
T
5
Press
to start the loss/length test.
6
To save the results, press
s
fiber ID; then press
Loopback Mode Test Results
The SUMMARY screen, which appears when the test is
finished, is described in Table 6-6.
To see more detailed results, press
Table 6-7 describes the RESULTS screen.
s
; select or create a
again.
A
View Results.