Using Manual OTDR Mode

You should use Auto OTDR mode when
certifying cabling with the OTDR.
Manual OTDR mode lets you optimize the OTDR trace for
viewing specific events.
To select Manual OTDR mode, exit to the HOME screen,
A
press
Change Test; then select Manual OTDR.
Setting
RANGE*
The range setting determines the maximum distance shown on the trace. Select the range that is
nearest to, but not less than, the distance to the event you want to study. If the tester does not
correctly identify the end event, run the OTDR test again using the next highest range.
Auto: the tester selects the maximum range that is valid with the RESOLUTION setting. If RESOLUTION
is set to Auto, the tester selects the range that is nearest to, but not less than the end of the fiber.
These ranges are not limited to the fixed ranges provided.
3-46
Note
Table 3-5. Manual OTDR Settings
Description and Selection Guidelines
Table 3-5 describes the settings you can change in Manual
OTDR mode and how the settings affect the trace. You
can access these settings on the OTDR tab in SETUP.
Tip: To see the settings used for an OTDR test, press
View Details on the OTDR SUMMARY screen; then press
A
OTDR Settings. The settings used may be slightly
different from the nominal values selected on the OTDR
tab.